Standard Incubators
600 Series
Operation Manual and Parts List
LT2068X1 Rev. 2 (7006841)
© 2010 Fisher Scientific Inc.
All rights reserved
Printed in U.S.A.
Isotemp

Questions and Answers

  • R
    Rhonda MathisAug 10, 2025
    What to do if my Fisher Scientific 6858 shows 'EEE' on the display?
    • D
      dana00Aug 10, 2025
      If the alarm LED is also lit, allow the incubator to cool to the set point. If the alarm LED is not on, check the thermocouple connections at the rear of the temperature controller. If the thermocouple is OK, replace the control.
  • L
    Loretta DownsAug 15, 2025
    How to fix a Fisher Scientific 6858 Laboratory Equipment that is not heating?
    • H
      Heather SmithAug 15, 2025
      First, ensure the set temperature is not less than the actual temperature. If not, replace the control thermocouple. Also, tighten connections at the terminal strip and/or heater. Check the heater resistance against the schematic and replace if necessary. Finally, replace the controller.
  • R
    Raymond RamirezAug 21, 2025
    What to do if my Fisher Scientific 6858 incubator temperature is erratically high?
    • K
      Kim ReeseAug 21, 2025
      If the incubator temperature is erratically high, the control thermocouple or the control board may be defective and require replacement.
  • N
    Nicholas JacobsonAug 27, 2025
    Why is the alarm LED on and the control higher than the set temperature on my Fisher Scientific 6858 Laboratory Equipment?
    • K
      Kimberly StoneAug 27, 2025
      This can happen if the set temperature has been changed to a value less than the actual temperature minus the alarm limit. Wait for the actual temperature to cool to the set temperature, specifically to set plus 3°, and the LED should go out and the EEE alarm display will return to normal operation. If this doesn't work, the controller or solid state relay/safety relay may be defective and require replacement.
  • R
    Raymond ColeSep 2, 2025
    How to troubleshoot a Fisher Scientific 6858 Laboratory Equipment with no power?
    • C
      Cheryl MartinezSep 2, 2025
      First, ensure the unit is plugged in and turned on. If it is, replace the circuit breaker if it's defective, or replace any loose wires.

Summary

Section 1 Introduction

Section 2 Specifications

Operating Range

Details the temperature range and performance characteristics of the incubator.

Electrical Requirements

Lists voltage, frequency, current, and wattage for different models.

Section 3 Installation

Unpacking

Instructions for unpacking the incubator and verifying contents.

Preparing the Incubator

Steps to prepare the incubator for operation after unpacking.

Power Switch

Details the operation and function of the front panel power switch.

Section 4 Controls

Display Fields

Describes the function of the LED displays for temperature, heat, alarm, and program status.

Section 5 Operation

Safety Precautions

Essential safety guidelines to follow before operating the incubator.

Alarm Limits

Explains the deviation alarm and how it functions when temperature limits are exceeded.

Section 6 Service

Section 7 Troubleshooting

Section 8 Replacement Parts

Section 10 Schematic

Fisher Scientific 6858 Specifications

General IconGeneral
BrandFisher Scientific
Model6858
CategoryLaboratory Equipment
LanguageEnglish