Standard Incubators
600 Series
Operation Manual and Parts List
LT2068X1 Rev. 2 (7006841)
© 2010 Fisher Scientific Inc.
All rights reserved
Printed in U.S.A.
Isotemp

Questions and Answers

  • J
    Joseph PattersonAug 9, 2025
    What to do if my Fisher Scientific 6841 shows 'EEE' on the display?
    • E
      Emily RamirezAug 9, 2025
      If the alarm LED is also lit, allow the incubator to cool to the set point. If the alarm LED is not on, check the thermocouple connections at the rear of the temperature controller. If the thermocouple is OK, replace the control.
  • J
    josephshannonAug 15, 2025
    How to fix a Fisher Scientific 6841 that is not heating?
    • R
      Rachel JonesAug 16, 2025
      First, ensure the set temperature is not less than the actual temperature. If not, replace the control thermocouple. Also, tighten connections at the terminal strip and/or heater. Check the heater resistance against the schematic and replace if necessary. Finally, replace the controller.
  • T
    Thomas PhillipsAug 22, 2025
    What to do if my Fisher Scientific 6841 Laboratory Equipment incubator temperature is erratically high?
    • F
      Frank CooperAug 22, 2025
      If the incubator temperature is erratically high, the control thermocouple or the control board may be defective and require replacement.
  • J
    James KingAug 27, 2025
    Why is the alarm LED on and the control higher than the set temperature on my Fisher Scientific 6841?
    • W
      Wendy RobinsonAug 27, 2025
      This can happen if the set temperature has been changed to a value less than the actual temperature minus the alarm limit. Wait for the actual temperature to cool to the set temperature, specifically to set plus 3°, and the LED should go out and the EEE alarm display will return to normal operation. If this doesn't work, the controller or solid state relay/safety relay may be defective and require replacement.
  • R
    reynoldschristopherSep 2, 2025
    How to troubleshoot a Fisher Scientific 6841 with no power?
    • L
      leejamieSep 3, 2025
      First, ensure the unit is plugged in and turned on. If it is, replace the circuit breaker if it's defective, or replace any loose wires.

Summary

Specifications

Installation

Setup Procedures

Details on preparing the incubator for operation, using the power switch, and understanding the convenience outlet.

Operation

Setting the Temperature

Step-by-step instructions for setting the desired temperature and initiating control mode.

Safety Precautions and Alarm Limits

Essential safety precautions for operating the incubator and details on the deviation alarm limits.

Service

Electronics Access and Heater Replacement

Instructions for accessing the electronics compartment and replacing the heater assembly.

Controller and Relay Replacement

Procedures for replacing the main controller and the solid-state relay.

Troubleshooting

Replacement Parts

Fisher Scientific 6841 Specifications

General IconGeneral
BrandFisher Scientific
Model6841
CategoryLaboratory Equipment
LanguageEnglish