AUTOMATIC STEAM STERILIZER

Questions and Answers

  • X
    Xavier WilliamsAug 9, 2025
    What to do if the low water indicator light is on and the alarm sounds on my Fisher Scientific Laboratory Equipment?
    • O
      Oscar LittleAug 9, 2025
      If the 'LOW WATER INDICATOR LIGHT' is on and an alarm sounds on your Fisher Scientific Laboratory Equipment, it could be due to several reasons. First, check if the water level is low and add water to the reservoir, ensuring it stays below the 'FULL' mark. If the issue persists, the water level sensor in the chamber might be unclean, so try cleaning it. Another potential cause is an unclean steam solenoid valve; in this case, clean the filter.
  • R
    Ryan RayAug 17, 2025
    What to do if water in the chamber can't return to the water reservoir automatically in my Fisher Scientific Laboratory Equipment?
    • D
      Deborah HallAug 18, 2025
      If water in the chamber of your Fisher Scientific Laboratory Equipment can’t return to the water reservoir automatically, the steam solenoid coil might be broken and needs to be replaced.
  • D
    Darlene HuertaAug 25, 2025
    How to fix a Fisher Scientific SterilElite16 Laboratory Equipment when the power light doesn't come on?
    • L
      Lance LeeAug 25, 2025
      If the POWER light indication doesn’t come on for your Fisher Scientific Laboratory Equipment, ensure the power supply is properly connected. If it still doesn't turn on, check the circuit breaker. If it's out, wait 3 minutes, then press it back in. If the circuit breaker can’t be reset, the unit requires servicing. Alternatively, the bulb or switch may be defective and need replacement.
  • Y
    yorozcoSep 1, 2025
    Why does the chamber pressure not increase in my Fisher Scientific SterilElite16?
    • J
      Jessica CruzSep 1, 2025
      If the chamber pressure does not increase in your Fisher Scientific Laboratory Equipment, start by cleaning the filter and solenoid valve. If that doesn't resolve the issue, the solenoid valve might need replacement. Another possibility is a defunct main heater, which would also require replacement. If the steam pressure is over exhaust or there are leaks in the steam piping system, call technical support.

Summary

CHAPTER 1. LOCATION

CHAPTER 2. TECHNICAL DETAILS

CHAPTER 3. PREPARATION & INSTALLATION

Unpacking Checklist

Steps for safely removing the autoclave from its packaging and ensuring all components are present.

First Time Use Checklist

Procedures for initial setup, including placement, airflow, and power checks before the first operation.

CHAPTER 4. CONTROL PANEL

CHAPTER 4 BASIC STEPS OF STERILIZATION

CHAPTER 4 OPERATION

Sterilization Box Best Practices

Key operational advice regarding the use and preparation of the sterilization box for optimal results.

CHAPTER 4. OPERATION - PROGRAMS

CHAPTER 5. REMINDERS, WARNINGS & CAUTION

CHAPTER 6. APPENDIX - A

CHAPTER 6. APPENDIX - B

CHAPTER 7. TROUBLE SHOOTING

CHAPTER 8. MAINTENANCE REQUIREMENTS

Warranty

Fisher Scientific SterilElite16 Specifications

General IconGeneral
BrandFisher Scientific
ModelSterilElite16
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals