Questions and Answers

  • M
    Maria MercadoAug 8, 2025
    What to do if calibration values are too far apart on Fisher Scientific Measuring Instruments?
    • I
      Isaiah MooreAug 9, 2025
      If the calibration values are too far apart, or too far from ideal values, check the electrode, use fresh buffers, and check the meter settings. The electrode may require filling solution, conditioning, or replacement. Refer to the electrode manual.
  • S
    Stacey GardnerAug 14, 2025
    How to calibrate temperature on Fisher Scientific Measuring Instruments?
    • S
      spierceAug 14, 2025
      Calibrate using a water bath or a known accurate thermometer. A faulty thermistor may also be the cause, consider repair or replacement.
  • A
    ashleycookAug 21, 2025
    What to do if Fisher Scientific Measuring Instruments show Over Range or Under Range?
    • J
      jeremycarsonAug 21, 2025
      Check that the electrode is connected.
  • M
    Michele MonroeAug 27, 2025
    Why are my pH calibration points too close on Fisher Scientific accumet AB250 Measuring Instruments?
    • X
      xmurphyAug 27, 2025
      Custom pH buffer calibration values must be at least 1.0 pH unit apart.
  • J
    Joseph LangSep 1, 2025
    What to do if 2-point standardization has not been performed on Fisher Scientific Measuring Instruments?
    • E
      Erik McneilSep 1, 2025
      Perform a 2-point standardization.