AIR FILTERING
The device is quipped with an extraction lter (A3) that must be replaced when
it is dirty or changes colour. Do not wash or reuse the same lter. Regular lter
replacement is necessary to help ensure proper compressor performance. The
lter must be checked regularly.
Contact your retailer or authorised service centre for replacement lters.
Remove the lter as shown in the picture to replace it.
The lter was designed to always remain secured in its seat. Do not replace the
lter during use.
Only use original accessories and spare parts by Flaem, we disclaim any
liability in the event of using non original spare parts or accessories.
After having sanitised the accessories, shake them vigorously and place them on a paper towel.
Alternatively, dry them with a jet of hot air (for example, a hair dryer).
DISINFECTION
After sanitising the nebuliser cup and the accessories, disinfect them choosing one of the methods
described below.
method A: Accessories C1.1-C1.2-C1.3-C1.4-C2-C3-C4 can be disinfected.
The disinfectant must be an electrolytic chlorine oxidiser (active principle: sodium hypochlorite) specic for
disinfecting, which is available in any pharmacy.
Directions:
- Fill a container of suitable size to hold all of the parts to be disinfected with a solution of potable water
and disinfectant, according to the proportions indicated on the packaging of the disinfectant.
- Completely immerse each part in the solution, taking care to avoid the formation of air bubbles on the
parts. Leave the parts immersed for the amount of time indicated on the packaging of the disinfectant
associated with the concentration chosen for the solution.
- Remove the disinfected parts and rinse with plenty of warm potable water.
- Dispose of the solution following the instructions provided by the disinfectant manufacturer.
method B: Sanitise the accessories C1.1-C1.2-C1.3-C1.4-C2-C3-C4 by boiling them in water for 10 minutes;
use demineralised or distilled water to prevent calcium deposits.
method C: Sanitise the accessories C1.1-C1.2-C1.3-C1.4-C2-C3-C4 with a hot steam steriliser for baby-
bottle (not the microwave type). Perform the process faithfully following the instructions of the steriliser.
To ensure that the disinfection is eective, choose a steriliser with an operating cycle of at least 6 minutes.
After having disinfected the accessories, shake them vigorously and place them on a paper towel.
Alternatively, dry them with a jet of hot air (for example, from a hair dryer).
At the end of each use store the device complete with accessories in a dry place away from dust.