EasyManua.ls Logo

Flaem ProfiNeb - Trouble-Shoting

Flaem ProfiNeb
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
17
TROUBLE-SHOTING
Switch o the device before any procedure and unplug the power cable from the socket.
PROBLEM CAUSE SOLUTION
The device does not work The power adaptor has not been
correctly inserted into the socket of
the device or the power socket
Correctly insert the power adaptor
in sockets.
The device does not
nebulize or nebulizes
insuciently.
The medication has not been
inserted in the nebulizer
Pour the right amount of medication
in the nebulizer
The nebulizer has not been prop-
erly tted.
Disassemble the nebulizer and reas-
semble it correctly as shown in the
connection diagram on the cover
The nebulizer nozzle is clogged Disassemble the nebulizer, remove
the nozzle and perform cleaning
operations Failure to remove medi-
cation deposits from the nebulizer
aects its eciency and operation.
Strictly comply with the instructions
contained in the CLEANING,
SANITISATION AND DISINFECTION
chapter
The air pipe is not correctly con-
nected to the device
Ensure the devices accessories
are properly connected to the air
intake (see connection diagram on
the cover).
The air pipe is bent, damaged or
twisted
Unwind the pipe and ensure it is
not crushed or punctured. If neces-
sary replace it.
The air lter is dirty Replace the lter
The apparatus is noisier
than usual
The Filter is not correctly
inserted in its housing
Fully and properly insert the Filter
in its housing
If after verifying the above mentioned conditions the device should not operate properly, contact your
trusted retailer or the nearest authorized FLAEM service center. You can nd the list of all the Service Cen-
tres at http://www.aemnuova.it/it/info/assistenza