16
Mod. P0915EM
Power source: 12V DC 1.25A
Max pressure: 2 ± 0.3 bar
Compressor air delivery: approx 9 l/min
Noise (at 1 m): approx 51 dB (A)
Operation: 30 minutes ON - 30 minutes OFF
Operating Conditions:
Temperature: min 10°C; max 40°C
Air humidity: min 10%; max 95%
Atmospheric pressure: min 69KPa; max 106KPa
Storage conditions:
Temperature: min -25°C; max 70°C
Air humidity: min 10%; max 95%
Atmospheric pressure: min 69KPa; max 106KPa
Dimensions: 12(L) x 12(P) x 6(H) cm
Weight: 0.58 Kg
Power supply
Only use power supply mod. UME318-1215 (cod. 17111)
Primary: 100/240V~ 50/60Hz
Secondary: 12V DC 1.5 A
APPLIED PARTS
Type BF applied parts are: patient accessories (C2, C3, C4)
RF7 Dual Speed Plus ampoule
Minimum medication capacity: 2 ml
Maximum medication capacity: 8 ml
Operating pressure (with neb.): 0.75 bar
TECHNICAL SPECIFICATIONS
C1.4 speed selector
Max Min With valve
(1)
Delivery: approx. 0.42 ml/min approx. 0.18 ml/min
(2)
MMAD: 2.44 μm 2.95 μm
(2)
Breathable fraction < 5 μm (FPF): 80.7 % 74.7 %
(1) data recorded according to internal Flaem procedure I29-P07.5.
(2) In vitro testing performed by TÜV Rheinland Italia S.r.l. in collaboration with the University of Parma and in compliance with the EN 13544-
1: 2007 + A1 European Standard for aerosol therapy devices Standard. More details are available on request.