Cooler maintenance
26
26.1 General
The microcooler is designed to provide maintenance-free operation for many thousands
of hours. The microcooler contains pressurized helium gas.
After several thousand hours of operation the gas pressure decreases, and cooler service
is required to restore cooler performance. The cooler also contains micro ball bearings,
which may exhibit wear by becoming louder.
26.2 Signs to watch for
The FLIR Systems microcooler is equipped with a closed-loop speed regulator, which ad-
justs the cooler motor speed to regulate the detector temperature.
Typically, the cooler runs at maximum speed for 7–10 minutes (depending on model), and
then slows to about 40% of maximum speed. As the gas pressure degrades, the motor
continues at maximum speed for longer and longer periods to attain operating
temperature
Eventually, as the helium pressure decreases, the motor will lose the ability to achieve
and/or maintain operating temperature. When this occurs, the camera must be returned to
FLIR Systems Customer Service Department for service.
#T559157; r. AH/45951/45951; en-US
231