FLIRTG54_TG56UserManual23DocumentIdentifier:TG54_TG56‐en‐US_AA
8.3IRThermometerSpecifications
IRTemp.Measurementrange‐22°F~+1202°F(‐30°C~+650°C)
Over‐andunder‐rangeindication OL
IRTemperatureResolution 0.1°C(0.1°Fupto999
o
Fand1
o
Fwhen≥ 1000
o
F)
IRTemperatureAccuracy ‐22°~+14°F(‐30°~‐10°C):±5.4°F(3.0°C)
+14°~+32°F(‐10°~<0°C):±3.6°F(2.0°C)
+32°~+1202°F(0°~+650°C):±1.8°F(±1.0°C)or±1.0%
(whicheverisgreater)
DistancetoSpot(D:S)ratio 24:1forTG54
30:1forTG56
Fieldofview
angle 0.04
o
IRTemperatureResponseTime 150ms
TemperatureScanning Continuous
Emissivity 4presetsplusacustomsetting(0.10~0.99)
8.4LaserSpecifications
Lasertype SingleClass1Laser(red;visiblelight)
Laserwavelength 650nm±20nm
Laserpower≤0.39mW
Laseralignment Paralleltomeasurement‘cone’;seeSection5.3,Distance
toSpotRatio
8.5ThermocoupleSpecifications
Thermocoupletype TypeK
Meterinputrange* ‐22°to1202°F(‐30°to650°C)
Meterinputaccuracy ±3.6°F(2°C)or±1%whicheverisgreater
Notincludingadditionalprobeerror:±4.5°F(2.5°C)
Measurementresolution 0.1°C(0.1°Fupto999
o
Fand1
o
Fwhen≥ 1000
o
F)
*IncludedTypeKproberange‐22°to572°F(‐30°to300°C)