Defibrillator/Transcutaneous Pacemaker Analyzer
Specifications
37
Asystole
Conduction
1° Block
2° Block Type I
2° Block Type II
3° Block
Right Bundle Branch Block RBBB
Left Bundle Branch Block LBBB
Transvenous Paced with selectable pacer spike amplitudes and widths
Atrial 80 BPM
Async 75 BPM
Demand with frequent sinus beats
Demand with occasional sinus beats
AV Sequential
Non-Capture
Non-Function
Selectable pacer pulse parameters for transvenous simulation. (Atrial and Ventricular channels are independently selectable):
Atrial Pacer Pulse
Width ....................................................0.1, 0.2, 0.5, 1.0, 2.0 ms
Polarity .................................................+ or -
Amplitude .............................................0 (off), 2 to 20 (by 2), 50, 100, 200, 500, 700 mV
Ventricular Pacer Pulse
Width ....................................................0.1, 0.2, 0.5, 1.0, 2.0 ms
Polarity .................................................+ or -
Amplitude .............................................0 (off), 2 to 20 (by 2), 50, 100, 200, 500, 700 mV