Defibrillator/Transcutaneous Pacemaker Analyzer
Using Pacemaker Test Elements (Impulse 7000DP only)
69
Table 17. ECG Pacer Interactive Test Custom Parameters
Parameter Description
Response Waveform The type of patient response to simulate:
Demand: Normal sinus rhythm at specified rate. Pacer in demand mode can interact so it
paces when the normal rate is too slow. Heart responds to pacer pulses at or above
threshold.
Asystole: No heartbeat, but heart responds to pacer pulses at or above threshold.
Non-capture: Same as Asystole, but heart fails to respond to one out of every ten pacer
pulses.
Non-function: No heartbeat and no response to pacing.
Rate Beats per minute to use for the Demand Response Waveform
Duration How long the simulation should last
Setting can range from 1 to 60 seconds. If the “Indefinite” checkbox is checked, the duration is
ignored, and the test must be manually stopped by the field user (Stop in TEST GUIDE toolbar).
Response Threshold The level of pacing current required to generate a heart response

Table of Contents

Other manuals for Fluke Biomedical IMPULSE 6000D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke Biomedical IMPULSE 6000D and is the answer not in the manual?

Fluke Biomedical IMPULSE 6000D Specifications

General IconGeneral
BrandFluke Biomedical
ModelIMPULSE 6000D
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals