14
In the command set below, the short form of parameters MINimum, MAXimum
and DEFault are used for brevity, the long form can be used if preferred.
CONFig:<meter_fn> <parameters>
:CONFigure:CAPacitance [<range>|MIN|MAX|DEF [,<resolution>|MIN|MAX|DEF]]
:CONFigure:CURRent:AC [<range>|MIN|MAX|DEF [,<resolution>|MIN|MAX|DEF]]
:CONFigure:CURRent:DC [<range>|MIN|MAX|DEF [,<resolution>|MIN|MAX|DEF]]
:CONFigure:XCURrent:AC [<range>|MIN|MAX|DEF [,<resolution>|MIN|MAX|DEF]]
:CONFigure:XCURrent:DC [<range>|MIN|MAX|DEF [,<resolution>|MIN|MAX|DEF]]
:CONFigure:VOLTage:AC [<range>|MIN|MAX|DEF [,<resolution>|MIN|MAX|DEF]]
:CONFigure:VOLTage:DC [<range>|MIN|MAX|DEF [,<resolution>|MIN|MAX|DEF]]
:CONFigure:RESistance [<range>|MIN|MAX|DEF [,<resolution>|MIN|MAX|DEF]]
:CONFigure:FRESistance [<range>|MIN|MAX|DEF [,<resolution>|MIN|MAX|DEF]]
:CONFigure:FREQuency [DEFault[,gate_time|MIN|MAX|DEF]]
:CONFigure:PERiod [DEFault[,gate_time|MIN|MAX|DEF]]
:CONFigure:POWer no parameters
:CONFigure:TEMPerature[:TRANsducer]:TCouple [<type>[,DEFault[,<resolution>|MIN|MAX|DEF]]]
:CONFigure:TEMPerature[:TRANsducer]:RTD [<DEFault >[,DEFault[,<resolution>|MIN|MAX|DEF]]]
:CONFigure:TEMPerature[:TRANsducer]:TRTD [<DEFault >[,DEFault[,<resolution>|MIN|MAX|DEF]]]
:CONFigure:TEMPerature[:TRANsducer]:FRTD [<DEFault >[,DEFault[,<resolution>|MIN|MAX|DEF]]]
CONFigure?
Returns the setup configured by the last CONFigure or
MEASure? See Note [6].
Notes for These Commands
Note 1:
CONFigure:<meter_fn> <parameters> forces the trigger system state to:
• INITiate:CONTinuous OFF
• ARM:LAYer2:SOURce IMMediate
• ARM:LAYer2:DELay 0 (ARM2:DELay:AUTo is set to OFF)
• ARM:LAYer2:COUNt 1
• ARM:LAYer1:SOURce IMMediate
• ARM:LAYer1:COUNt 1
• ARM:LAYer1:DELay 0 (ARM2:DELay:AUTo is set to OFF)
• TRIGger:SOURce IMMediate
• TRIGger:COUNt 1
• TRIGger:DELay 0.0 (TRIGger:DELAY:AUTO is set to OFF)
• TRIGger:HOLDoff AUTO