8588A/8558A
Remote Programmer's Manual
50
SCPI QUEStionable Status Register
Bit No. Bit Value SCPI Defined 8588A Meaning
0 1 VOLTage Over-range
1 2 CURRent Over-range
2 4 TIME Not Used
3 8 POWER Over-range (when RF sensor fitted)
4 16 TEMPerature Over-range (PRT and Thermocouple)
5 32 FREQuency Over-range (Counter)
6 64 PHASe Not Used
7 128 MODulation Not Used
8 256 CALibration Not Used
9 512 Available to Designer Under range (All functions)
10 1024 Available to Designer Over range (All functions)
11 2048 Available to Designer Trigger too fast
12 4096 Available to Designer Measurement too fast
13 8192 INSTrument Summary Not Used
14 16384 Command Warning Not Used
15 32768 NOT USED