SCPI OPERation Status Register
Bit No. Bit Value SCPI Defined 8588A Meaning
0 1 CALibrating Not Used
1 2 SETTing Not Used
2 4 RANGing Not Used
3 8 SWEeping Not Used
4 16 MEASuring STATus:CONDition is false when the
Trigger system in IDLE state, true when
not IDLE
STATUS:EVENt is true when one or
more readings are available
5 32 Waiting for TRIGger Waiting for TRIGger
6 64 Waiting for ARM Waiting for ARM
7 128 CORRecting Not Used
8 256 Available to Designer Low Limit
9 512 Available to Designer High Limit
10 1024 Available to Designer Ext. Ref. clock unlock
11 2048 Available to Designer Signal clipping (Digitize & AC)
12 4096 Available to Designer Not Used
13 8192 INSTrument Summary Not Used
14 16384 PROGram Running Not Used