Reference Multimeter and 8 ½ Digit Multimeter
Remote Command Examples
59
Remote Command Examples
On power up or after the *RST reset command, the DMM is in this state:
Function: DCV
Range: 1000V
Resl:1.000E-04 (100 µV)
Aperture:/NPLC 50 PLC
IMP: Auto (10 MΩ while in the 1000 V range)
Remote measurement response format: ASCII
Result bit order = Swapped
MATH and LIMIT checking OFF
Memory Location: Volatile Buffer only
ROUTe:TERMinals Front
ROUT:TOUTput = Signal Acquired
ROUTe:INPut:GUARd = OFF
CALCulate:AVERage = 1, OFF
Product settings:
Display: Unchanged
Line Frequency: Auto
Ext Ref Clock: Off
Remote settings:
Emulation: Unchanged
Active Port: Unchanged
GPIB Address: Unchanged
Ethernet settings: Unchanged