Reference Multimeter and 8 ½ Digit Multimeter
SCPI Commands Reference
37
Table 6. Command Summary (cont.)
Keyword Description
:RANGe:AUTO[:STATe]<boolean> Turns Auto range ON or OFF
:RANGe:AUTO[:STATe]? Returns 1 for Auto range ON, 0 for auto range OFF
:RANGe < nrf >|MIN|MAX|DEF
Set maximum expected value or MIN, MAX, or
Default range
:RANGe? [MIN|MAX|DEF]
Returns the selected range or, or if specified, the
MIN, MAX, or Default range
:RESolution < nrf >|MIN|MAX|DEF
Set maximum expected value or MIN, MAX, or
Default resolution; for example, range is 1 A,
Resolution <nrf>. = 0.0001 (100 µA), the
measurement is returned with a resolution of
+1.000E-4
:RESolution? [MIN|MAX|DEF]
Returns the selected resolution, or if specified, the
MIN, MAX, or Default range
:SECondary
OFF|CFACtor|FREQuency|NPEak|PERiod
|PTP|PPEak|PUNCertainty|VOLTage
Sets the secondary reading type
OFF = Secondary reading is not shown
CFACtor = Secondary reading is the crest factor
FREQuency = Secondary reading is frequency
NPEak = Secondary reading is negative peak
PERiod = Secondary reading is period
PTP = Secondary reading is peak-to-peak
PPEak = Secondary reading is positive peak
PUNCertainty = additional Power Uncertainty
VOLTage = the shunt output voltage
:SECondary? Returns the selected secondary reading setting