Reference Multimeter and 8 ½ Digit Multimeter
SCPI Commands Reference
39
Table 6. Command Summary (cont.)
Keyword Description
:SHUNt:LIST? Returns a comma-separated enumerated list of
shunts as {List#, Asset#, Manufacturer, Model,
Serial#, Resistance, Max Current, Power Ref Level,
Power Coefficient}
Note1: The first four items are quoted strings, the
remaining item are NR3 format values.
Note2: If the optional <list#> is supplied that record is
returned, otherwise the active shunt record is
returned.
Note3: Index ‘0’ is the ‘Basic' shunt slot, and is always
available.
SHUNt:MODify <list#>,< list> <list> See notes to SHUNt?
:SHUNt:PICK <list#> selects shunt by enumerated list number
:SHUNt:PICK? Returns selected shunt number
:SHUNt:REMove <list#> Remove the shunt at # in the list
SHUNt:SORT MCURrent|ASSet|SERial
Sort shunt list in either ‘maximum current’, ‘asset
number’ or ‘serial number’ order
SHUNt:SORT? Returns the sort order (MCUR|ASS|SER)
Note(1): External shunts data is common between SENSe:XCURrent:AC and SENSe:XCURrent:DC
[:SENSe]:XCURrent:DC
:APERture <nrf>|MIN|MAX|DEF Sets the ADC aperture value in seconds or to the
MIN, MAX, or Default setting
:APERture? [MIN|MAX|DEF] Returns the aperture setting or the MIN, MAX, or
Default settings in seconds
:APERture:MODE AUTO|FAST|MANual Sets the aperture mode
:APERture:MODE? Returns the aperture mode
:NPLC <nrf>MIN|MAX|DEF Sets the ADC aperture in number of power line cycles
or to the MIN, MAX, or Default plc setting
:NPLC? [MIN|MAX|DEF] Returns the ADC aperture in number of power line
cycles or to the MIN, MAX, or Default plc setting
:RANGe:AUTO[:STATe]<boolean> Turns Auto range ON or OFF
:RANGe:AUTO[:STATe]? Returns 1 for Auto range ON, 0 for auto range OFF
:RANGe <nrf>|MIN|MAX|DEF Set maximum expected value or MIN, MAX, or Default
range