8588A/8558A
Remote Programmer's Manual
52
Table 11. SYSTem Subsystem (cont.)
Keyword Description
:TEMPerature? Returns internal temperatures in form x,y
x = temperature in degrees Celsius measured on the
analogue assembly
x = temperature in degrees Celsius measured on the Digital
assembly
:COMMunicate:ENABLE
USB|GPIB | ENET
Selects active remote port (no query available for this
command)
:COMMunicate:GPIB:ADDress
<value>
Set GPIB address
:COMMunicate:GPIB:ADDress? Query the GPIB bus address
:COMMunicate:LAN:CONTrol <port> Set or query the port
:COMMunicate:LAN:CONTrol?
:COMMunicate:LAN:DHCP[:STATE]
<boolean>
Turn DHCP ON or OFF
:COMMunicate:LAN:DHCP[:STATE]?
:COMMunicate:LAN:GATeway
<ipaddress>
<ipaddress> is in the form xxx.xxx.xxx.xxx, i.e.
192.168.1.190
:COMMunicate:LAN:GATeway? Returns the address
:COMMunicate:LAN:IPADdress
<ipaddress>
:COMMunicate:LAN:IPADdress?
:COMMunicate:LAN:MAC? Returns the MAC address
:COMMunicate:LAN:SMASk
<ipmask>
<ipmask> is in the form xxx.xxx.xxx.xxx, i.e. 255.255.255.0
:COMMunicate:LAN:SMASk? Returns the IP mask.
Note (1): Only a valid command when the Active Port is Ethernet. The commands cause an error if the
Active Port is GPIB or USB.