True-rms Multimeters 
 General Specifications  
 21 
Table 1 Accuracy Specifications (cont.) 
Function Range Resolution 
Accuracy 
± ([% of Reading] + [Counts]) 
Model 
Continuity 600 Ω 1 Ω 
Beeper on < 20 Ω, off > 250 Ω; 
detects opens or shorts of  
500 μs or longer. 
115C, 
117C 
Ohms 
600.0 Ω 
6.000 kΩ 
60.00 kΩ 
600.0 kΩ 
6.000 MΩ 
40.00 MΩ 
0.1 Ω 
0.001 kΩ 
0.01 kΩ 
0.1 kΩ 
0.001 MΩ 
0.01 MΩ 
0.9 % + 2 
0.9 % + 1 
0.9 % + 1 
0.9 % + 1 
0.9 % + 1 
5 % + 2 
115C, 
117C 
Diode test  2.000 V  0.001 V 
0.9 % + 2 
115C, 
117C 
Capacitance  1000 nF 
10.00 μF 
100.0 μF 
9999 μF
 
1 nF 
0.01 μF 
0.1 μF 
1 μF
 
1.9 % + 2 
1.9 % + 2 
1.9 % + 2 
100 μF - 1000 μF: 1.9 % + 2 
> 1000 μF: 5 % + 20
 
115C, 
117C 
Lo-Z Capacitance 
(Power-up option) 
1 nF to 500 μF  10% + 2 typical 
115C, 
117C