1
58
7
/
1577 Insulation Multimeters Fluke Corporation 5
Model 1587
Output Voltage Display Range Resolution Test Current Resistance Accuracy
50 V
0.01 to 6.00 M 0.01 M
1 mA @ 50 k 3 % + 5 counts
(0 % to + 20 %)
6.0 to 50.0 M 0.1 M
100 V
0.01 to 6.00 M 0.01 M
(0 % to + 20 %)
6.0 to 60.0 M 0.1 M 1 mA @ 100 k 3 % + 5 counts
60 to 100 M 1 M
250 V
0.1 to 60.0 M 0.1 M
1 mA @ 250 k 1.5 % + 5 counts
(0 % to + 20 %)
60 to 250 M 1 M
500 V
0.1 to 60.0 M 0.1 M
1 mA @ 500 k 1.5 % + 5 counts
(0 % to + 20 %)
60 to 500 M 1 M
1000 V
0.1 to 60.0 M 0.1 M
1.5 % + 5 counts
(0 % to + 20 %)
60 to 600 M
1 M 1 mA @ 1 M
0.6 to 2.0 G 100 M 10 % + 3 counts
Model 1577
Output Voltage Display Range Resolution Test Current Resistance Accuracy
500 V
0.1 to 60.0 M 0.1 M 1 mA @ 500 k 2.0 % + 5 counts
(0 % to + 20 %)
60 to 500 M 1 M
1000 V
0.1 to 60.0 M 0.1 M 1 mA @ 1 M 2.0 % + 5 counts
(0 % to + 20 %)
60 to 600 M 1 M
Maximum voltage applied to any terminal: 1000 V ac rms
or dc
Storage temperature: -40 °C to 60 °C (-40 °F to 140 °F)
Operating temperature: -20 °C to 55 °C (-4 °F to 131 °F)
T
emperature coefficient:
0.05 x (spec
ified ac
curac
y) per
°
C
for temperatures < 1
8
°
C or > 28
°C (< 64 °F or > 82 °F)
Relative humidity, non-condensing: < °C
0 % to 95 % @ 10 °C to 30 °C (50 °F to 86 °F)
0 % to 75 % @ 30 °C to 40 °C (86 °F to 104 °F)
0 % to 40 % @ 40 °C to 55 °C (104 °F to 131 °F)
Vibration: Random, 2 g, 5-500 Hz per M
I
L-PRF-28800F,
Class 2 instrument
Shock: 1 meter drop per IEC 61010-1 2
nd
Edition (1 meter
drop test, six sides, oak floor)
Electromagnetic compatibility
:
In an R
F field of 3 V
/
M,
ac
curacy = specified accuracy except in temperature: specified
accuracy ± 5 °C (9 °F). (EN 61326-1:1997)
Safety: Complies with ANSI/ISA 82.02.01 (61010-1) 2004,
CAN/CSA-C22.2 NO. 61010-1-04, and IEC/EN 61010-1 2
nd
Edition for measurement C
AT III 1000 V and CAT IV 600 V
Certifications: CSA per standard CSA/CAN C22.2
No. 61010.1-04; TUV per standard EN 61010 Part 1-1002
Batteries: Four AA batteries (NEDA 15A or IEC LR6)
Battery life: Meter use 1000 hours; Insulation test use: Meter
can perform at least 1
000 insulation tests w
ith fresh alkaline
batteries at room temperature. These are standard tests of
1000 V into 1 M
with a duty cycle of 5 seconds on and 25
seconds off
Size: 5.0 cm H x 10.0 cm W x 20.3 cm L (1.97 in H x 3.94 in
W x 8.00 in L)
W
eight:
550 g (
1
.2 lb)
IP rating: IP40
Altitude (operating): 2000 m CAT III 1000 V, CAT IV 600 V;
3000 m C
A
T I
I 1
000 V, C
A
T III 600 V
Storage: 1
2,000 m
Over-
range capability:
1
10 % of range except for
capacitance which is 1 %
Compliance to EN 61557: IEC61557-1, IEC61557-2
1587/1577 General Specifications