EasyManua.ls Logo

Fluke 1620A DEWK - Specifications and Environmental Conditions; Specifications

Fluke 1620A DEWK
25 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7
2 Specifications and Environmental Conditions
Specifications and Environmental Conditions
Specifications
Temperature Range
0 °C to 50 °C
Temperature Accuracy
(“H” Model)
16 °C to 24 °C (60.8 °F to 75.2 °F): ± 0.125 °C (± 0.225 °F) [calibrated]
0 °C to 16 °C (32 °F to 60.8 °F): ± 0.5 °C (± 0.9 °F) [uncalibrated typical]
24 °C to 50 °C (75.2 °F to 122 °F): ± 0.5 °C (± 0.9 °F) [uncalibrated typical]
Temperature Accuracy
(“S” Model)
15 °C to 35 °C (59 °F to 95 °F): ± 0.25 °C (± 0.45 °F) [calibrated]
0 °C to 15 °C (32 °F to 59 °F): ± 0.5 °C (± 0.9 °F) [uncalibrated typical]
35 °C to 50 °C (95 °F to 122 °F): ± 0.5 °C (± 0.9 °F) [uncalibrated typical]
Delta Temperature
Accuracy
± 0.025 °C (± 0.045 °F) for ± 1 °C (± 1.8 °F)
changes within 15 °C to 35 °C (59 °F to 95 °F)
Temperature Display
Resolution
User selectable up to 0.001 °C (0.01 °C recorded)
RH Range
0 % to 100 %RH
RH Accuracy
(“H” Model)
20 % to 70 %RH: ± 1.5 %RH (calibrated)
0 % to 20 %RH, 70% to 100 %RH: ± 3 %RH (uncalibrated, typical)
RH Accuracy
(“S” Model)
20 % to 70 %RH: ± 2 %RH (calibrated)
0 % to 20 %RH, 70% to 100 %RH: ± 3 %RH (uncalibrated, typical)
Delta Humidity
Accuracy
± 1.0% for ± 5% changes within 20% to 70 %RH
RH Display Resolution
User selectable up to 0.01% (0.1% recorded)
Inputs
Two sensors, each measuring temperature and relative humidity; each is
detachable, cable-extendable, and interchangeable, with self-contained
calibration; each may be assigned a unique 16-character identification
Display
240 x 128 graphics monochrome LCD, displays temperature and humidity
data graphically, numerically, and statistically; 16 pre-defined, user-
changeable screen set-ups are included
Memory
400,000 typical individual time-stamped readings
Alarms
Visual and audio alarms for temperature, temperature rate, RH, RH rate,
and fault conditions
Alarm port output
0 V normal, 11 to 12 V active, sources up to 20mA, 2.5mm, two-conductor
subminiature plug
Communications
RS-232, Ethernet LAN, 802.15.4 (ZigBee) wireless (optional)
Ethernet
10 Base-T, 100 Base-TX, IP, TCP, DHCP, Ping, HTTP, HTML
Wireless Range
30 m (100 ft) typical unobstructed
Enclosure
The DewK may be wall-mounted (hardware included) or set on a benchtop
Power
12 V DC from external 100-240 V AC power supply
Battery Backup
Standard 9V battery to allow continued measuring during power disruptions
Operating Range
0 °C to 50 °C
Size (DewK)HxWxD
125 mm x 211 mm x 51 mm (4.9 in x 8.3 in x 2.0 in)
Size (Probes)
79 mm H x 19 mm dia. (3.1in x 0.75 in)
Weight
0.7 kg (1.5 lb.)