Earth/Ground Tester
Setup
7
Operating Instructions
R
A
2-pole, 3-pole Measurements
To make 2-pole or dead-earth measurements, connect a jumper between
terminals H/C2 and S/P2 with the supplied connector cable. Refer to Figures 3
and 4. Then follow steps 1 thru 4 except use only earth electrode and the
auxiliary earth electrode with minimum distance between them of 20 m.
A Select function R
A
3-pole.
Display is as shown below.
B Connect test leads
Connect terminal E/C1 to the earth/ground system to be measured
with the supplied test lead and clip (1.5 m). Place 2 ground stakes in
earth/dirt. Minimum distance between earth electrode (E/C1), probe
(S/P2), and auxiliary earth (H/C2) should be at least 20 m!
Connect the stakes with the 25 m and 50 m cable reels to H/C2 and
S/P2 as shown below.
C Press START.
The “active” symbol indicates that a measurement is in progress. For
a continuous measurement keep START pressed.
D The Symbol 9” indicates a completed measurement. The result is
kept on the display until a new measurement is started or the main
switch is turned.
Shop for Fluke products online at:
1.877.766.5412
www.MyFlukeStore.com

Other manuals for Fluke 1623

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 1623 and is the answer not in the manual?

Fluke 1623 Specifications

General IconGeneral
2-Pole MeasurementYes
3-Pole MeasurementYes
4-Pole MeasurementYes
Data TransferUSB
Measurement Frequency128 Hz
Resistance Measurement Range0.01 Ω to 20 kΩ
Safety ComplianceIEC 61010-1
DisplayDigital
Power Supply6 x AA alkaline batteries
Operating Temperature0 °C to 50 °C (32 °F to 122 °F)
Storage Temperature-20 °C to 60 °C (-4 °F to 140 °F)