EasyManua.ls Logo

Fluke 187 - Page 93

Fluke 187
96 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specifications
Detailed Accuracy Specifications
7
7-9
Function Ranges Resolution Accuracy
Capacitance
2
1.000 nF 0.001 nF 2% + 5
10.00 nF 0.01 nF 1 % + 5
100.0 nF 0.1 nF
1.000 µF 0.001 µF
10.00 µF 0.01 µF
100.0 µF 0.1 µF
1,000 µF 1 µF
10.0 mF 0.01 mF
50.00 mF 0.01 mF
3
3 % + 10
Diode Test
1
3.1000 V 0.0001 V 2 % + 20
1. For the 5,000 count mode, divide the number of least significant digits (counts) by 10.
2. For film capacitor or better, using Relative mode (REL ) to zero residual on 1.000 nF and 10.00 nF ranges.
3. Least significant digit not active above 10 mF.