Specifications
23
Function Range Resolution Accuracy Typical Burden Voltage
32.00 mA 0.01 mA ±(0.75%+2) 5.6 mV/mA
A 320.0 mA 0.1 mA ±(0.75%+2) 5.6 mV/mA
10.00 A 0.01 A ±(0.75%+2) 50 mV/A
M
320.0 µA 0.1 µA ±(0.75%+2) 0.5 mV/µA
3200 µA 1 µA ±(0.75%+2) 0.5 mV/µA
E
32.00 mA 0.01 mA ±(1.5%+2) 5.6 mV/mA
40-1000 Hz 320.0 mA 0.1 mA ±(1.5%+2) 5.6 mV/mA
10.00 A 0.01 A ±(1.5%+2) 50 mV/A
N
320.0 µA 0.1 µA ±(1.5%+2) 0.5 mV/µA
40-1000 Hz 3200 µA 1 µA ±(1.5%+2) 0.5 mV/µA
* Accuracy is specified as ±([% of reading] + [number of least significant digits]).
Basic electrical accuracy is specified from 18°C to 28°C with relative humidity up to 95%, for a period of one year after
calibration. All ac conversions are ac coupled, average responding, and calibrated to read the true rms value of a sine wave
input.
Ranging is either automatic or manual in all functions with more than one range. Test resistance below approximately 270 e
in the Diode test function produces a continuous audible tone.