Specifications
25
Digital Display ......................................................................
3200 counts, updates 2/sec
Analog Display ..................................................................... 31 Segments, updates 25/sec
Operating Temperature ....................................................... -15°C to 55°C, to -40°C for 20 minutes when taken from
20°C
Storage Temperature
Without Battery .................................................................
With Battery ......................................................................
-55°C to 85°C
-55°C to 60°C
Electromagnetic Compatibility ........................................... In an RF field of 3 V/m on all ranges and functions except
mVAC, total accuracy = specified accuracy +1.0% of range.
For mVAC, total accuracy = specified accuracy + 1.5% of
range. EN 61326-1:1997
Temperature Coefficient
......................................................
0.1X (specified accuracy)/ °C (<18°C or >28°C)
Relative Humidity ................................................................. 0% to 95% (0°C to 35°C)
0% to 70% (35°C to 55°C)
Battery Type ......................................................................... 9 V, NEDA 1604 or 6F22 or 006P
Battery Life ........................................................................... 1000 hrs. typical
Shock, Vibration and Water Resistance ............................ Per MIL-T-28800 for a Style A, Class 2 Instrument
Size (H X W X L ) .................................................................. 2.2 in X 3.75 in X 8 in (5.6 cm X 9.5 cm X 20.3 cm)
Weight ................................................................................... 1.6 pounds (0.75 kg)
Altitude .................................................................................
Safety ....................................................................................
2000 m maximum
Complies with ANSI/ISA S82.01-1994, CAN/CSA C22.2 No.
1010.1:1992, 1000 V Overvoltage Category III. UL license
pending to UL3111-1.