EasyManuals Logo

Fluke 287 Specifications

Fluke 287
2 pages
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #2 background imageLoading...
Page #2 background image
Specifications
Function Range and Resolution Basic Accuracy
DC volts
AC volts
50.000 mV,500.00 mV, 5.0000 V, 50.000 V,
500.00 V, 1000.0 V
0.025 %
0.4 % (true-rms)
DC current
AC current
500.00 µA, 5000.0 µA, 50.000 mA, 400.00 mA,
5.0000 A, 10.000 A
0.15 %
0.7 % (true-rms)
Temperature
(excluding probe)
-200.0 °C to 1090.0 °C (-328.0 °F to 1994.0 °F) 1.0 %
Resistance
500.00 Ω, 5.0000 kΩ, 50.000 kΩ, 500.00 kΩ, 5.0000
MΩ, 50.00 MΩ, 500.0 MΩ
0.05 %
Capacitance 1.000 nF,10.00 nF 100.0 nF, 1.000 µF, 10.00 µF,
100.0 µF, 1000 µF, 10.00 mF, 100 mF
1.0 %
Frequency 99.999 Hz, 999.99 Hz, 9.9999 kHz, 99.999 kHz,
999.99 kHz
0.005 %
Ordering
information
287 True-rms Electronics
Logging Multimeter
with TrendCapture
Optional accessories
FVF-SC2 FlukeView
®
Forms
Software with Cable
80BK Integrated DMM
Temperature Probe
TLK287 Electronic Test Lead Set
TPAK Magnetic Hanging Kit
C280 Soft Case
Additional Functions/Features Fluke 287
Multiple on screen displays Yes
True-rms ac bandwidth 100 kHz
dBV/dBm Yes
DC mV resolution 1 µV
Megohm range up to 500 M
Conductance 50.00 nS
Continuity beeper Yes
Battery/fuse access Battery/fuse
Elapse time clock Yes
Time of day clock Yes
Min-max-avg Yes
Peak 250 µS
Duty cycle 0.01 % to 99.99 %
Pulse width 0.025 ms, 0.25 ms, 2.5 ms, 1250.0 ms
Hold Yes
Isolated optical interface Yes
Auto/touch hold Yes
Reading memory Yes
Log to PC Yes
Interval/event logging Yes
Logging memory up to 10,000 readings
Fluke Corporation
PO Box 9090, Everett, WA USA 98206
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or
Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2007 Fluke Corporation. All rights reserved.
Specifications subject to change without notice.
Printed in U.S.A. 5/2007 3032814 D-EN-N Rev A
Fluke. Keeping your world
up and running.
®
2 Fluke Corporation 287 True-rms Electronic Logging Multimeter with TrendCapture
General
specifications
Maximum voltage between any
terminal and earth ground:
1000 V
Battery type: 6 AA alkaline bat-
teries, NEDA 15A IEC LR6
Battery life: 50 hours minimum,
180 hours in logging mode
Temperature:
Operating: -20 °C to 55 °C
Storage: -40 °C to 60 °C
Relative humidity: 0 to 90 %
(0 t 37 °C), 0 to 65 % (37 °C to
45 °C), 0 to 45 % (45 °C to 55 °C)
Electromagnetic compatibility:
EMC EN61326-1
Vibration: Random vibration per
MIL-PRF-28800F Class 2
Shock: 1 meter drop per IEC/EN
61010-1 2nd Edition
Size (HxWxL): 22.2 cm x 10.2
cm x 6 cm (8.75 in x 4.03 in x
2.38 in)
Weight: 870.9 g (28 oz)

Other manuals for Fluke 287

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 287 and is the answer not in the manual?

Fluke 287 Specifications

General IconGeneral
Data holdYes
Product typeDigital multimeter
Product colorGrey, Yellow
Frequency range0 - 999.99 kHz
DC current range0 - 10 A
DC voltage range0 - 1000 V
Resistance range0 - 500000000 Ω
Capacitance range0 - 100000 nF
Resolution (AC voltage)0.001 mV
Basic accuracy (frequency)0.005%
Basic accuracy (DC current)0.05%
Basic accuracy (DC voltage)0.025
Temperature measurement range-200 - 1350 °C
Battery typeAA
Power sourceBattery
Battery technologyAlkaline
Number of batteries supported6
Storage temperature (T-T)-40 - 60 °C
Operating temperature (T-T)-20 - 55 °C
Operating relative humidity (H-H)0 - 90 %
Display digits4.5 digits
Weight and Dimensions IconWeight and Dimensions
Depth60 mm
Width102 mm
Height222 mm
Weight870.9 g

Related product manuals