Performance Tests
13
Table 4. Performance Tests
Test
(Switch
Position)
Calibrator
Output
323 324 325
Meter Reading Limit
Low High
Temperature
-5 °C N/A X X -5.8 °C -4.2 °C
0 °C N/A X X -0.8 °C 0.8 °C
100 °C N/A X X 98.2 °C 101.8 °C
400 °C N/A X X 395.2 °C 404.8 °C
AC Volts
10 V @ 50 Hz X X X 9.3 V 10.7 V
500 V @ 50 Hz N/A X X 492.0V 508.0 V
500 V @ 500 Hz N/A X X 492.0V 508.0 V
600 V @ 50 Hz X N/A N/A 590.5 V 609.5 V
600 V @ 400 Hz X N/A N/A 590.5 V 609.5 V
DC Volts
-600 V X N/A N/A -606.5 -593.5 V
-500 V N/A X X -505.5 V -494.5 V
10 V X X X 9.4 V 10.6 V
500 V N/A X X 494.5 V 505.5 V
600 V X N/A N/A 593.5 V 606.5 V
Ohms
0 Ω X N/A N/A 0.0 Ω 0.5 Ω
10 Ω X X X 9.4 Ω 10.6 Ω
300 Ω N/A X X 296.5 Ω 303.5 Ω
350 Ω X N/A N/A 346.0 Ω 354.0 Ω
1000 Ω N/A X X 985 Ω 1015 Ω
3000 Ω N/A X X 2965 Ω 3035 Ω
3500 Ω X N/A N/A 3460 Ω 3540 Ω
10K Ω N/A X X 9.85 KΩ 10.15 KΩ
30K Ω N/A X X 34.65 KΩ 30.35 Ω
Capacitance
10 μF N/A X X 9.5 μF 10.5 μF
500 μF N/A X X 491 μF 509 μF
900 μF N/A X X 887 μF 913 μF
Frequency
0.2 A, 5 Hz N/A N/A X 4.7 Hz 5.3 Hz
0.2 A, 50 Hz N/A N/A X 49.6 Hz 50.4 Hz
0.2 A, 500 Hz N/A N/A X 497.5 Hz 501.5 Hz