Loading...
323/324/325
Calibration Manual
14
Table 4. Performance Tests (cont.)
Test
(Switch
Position)
Calibrator
Output
323 324 325
Meter Reading Limit
Low High
AC Amps
(with 50-
Turn Coil)
0.12 A @ 50 Hz X N/A N/A 5.4 A 6.6 A
0.48 A @ 150 Hz X N/A N/A 22.9 A 25.1A
4.2 A @ 50Hz X N/A N/A 205.3 A 214.7 A
7.8 A @ 50 Hz X N/A N/A 381.7 A 398.3 A
7.8 A @ 150 Hz X N/A N/A 379.7 A 400.3 A
7.8 A @ 400 Hz X N/A N/A 379.7 A 400.3 A
0.004 A @ 50 Hz N/A X N/A 0.15 A 0.25 A
0.2 A @ 50 Hz N/A X N/A 9.80 A 10.20 A
0.6 A @ 50 Hz N/A X N/A 29.50 A 30.50 A
2 A @ 50 Hz N/A X N/A 97.5 A 102.5 A
6 A @ 50 Hz N/A X N/A 293.5 A 306.5 A
0.2 A @ 50 Hz N/A N/A X 9.75 A 10.25 A
0.6 A @ 50 Hz N/A N/A X 29.35 A 30.65 A
2 A @ 50 Hz N/A N/A X 98.0 A 102.0 A
6 A @ 50 Hz N/A N/A X 295.0A 305.0 A
DC Amps
(with 50-
Turn Coil)
0.2 A N/A N/A X 9.75 A 10.25 A
0.6 A N/A N/A X 29.35 A 30.65 A
2 A N/A N/A X 97.5 A 102.5 A
6 A N/A N/A X 293.5 A 306.5 A