365
Calibration Manual
10
Table 4. Adjustment Procedure (cont.)
Test
(Switch
Position)
Product
LCD
Reading
Action
C012 Calibrator output 10A @ 50 HZ, allow the reading to stabilize, wait 4
seconds, Push
C013 Calibrator output 2.5A @ 50 HZ, allow the reading to stabilize, wait 4
seconds, Push
C014 Calibrator output 2.5A @ 100 HZ, allow the reading to stabilize, wait 4
seconds, Push
C015 Calibrator output 2.5A @ 200 HZ, allow the reading to stabilize, wait 4
seconds, Push
C016 Calibrator output 2.5A @ 300 HZ, allow the reading to stabilize, wait 4
seconds, Push
C017 Calibrator output 2.5A @ 400 HZ, allow the reading to stabilize, wait 4
seconds, Push
?
AC Amps
Save Push
C018 Calibrator output 0A, allow the reading to stabilize, wait 4 seconds,
Push
C019 Calibrator output 10 A, allow the reading to stabilize, wait 4 seconds,
Push
A
DC Amps
Save Push