Detachable Jaw True-rms Clamp Meter
Safety Information
3
• Use caution around bare conductors or bus bars. To prevent
electrical shock, do not touch the conductor.
• Compl
y with local and national safety codes.
Use personal
protective equipment (approved rubber gloves, face
protection, and flame-resistant clothes) to prevent shock
and arc blast injury
w
here hazardous live conductors are
exposed.
• Disconnect po
we
r and discharge all high-voltage capacitors
before
you measure resistance or continuity.
• Do not measure ac/dc current in circuits carry
ing more than
600 V or 200 A with the Product Jaw.
• Do not operate the product with covers removed or the case
open. Hazardous voltage exposure is possible.
• When
white wear indicator insulation shows th
rough the
clamp cable jacket, replace the clamp cable.
• When batteries are changed, ensure that the calibration seal
in the battery compartment is not damaged. If damaged, the
Product may not be safe for use. Return the Product to
Fluke for replacement of the seal.
• Do not exceed the Measurement Category
(CAT) rating
of
the lo
w
est rated individual component of a product, probe,
or accessory
.
• Measure a known volt
age first to make sure that the product
operates correctl
y
.
W Caution
To prevent possible damage to the product or to equipment
under test:
• Use the correct termin
als, function, and range for
measurements.
• Clean the case and accessories w
ith a dam
p cloth and mild
detergent onl
y. Do not
use abrasives or solvents.