374/375/376
Calibration Manual
14
Table 4. Adjustment Procedure (cont.)
Test
(Switch Position)
LCD
Reading
374 375 376 Calibrator Output
C-23 X X X 8 A @ 50 Hz
C-24 X X X 3 A @ 50 Hz
C-25 X X X 3 A @ 100 Hz
C-26 X X X 3 A @ 200 Hz
C-27 X X X 3 A @ 300 Hz
C-28 X X X 3 A @ 400 Hz
C-29 X X X 3 A @ 440 Hz
?
AC Amps
(with 50-turn Coil)
Save X X X STBY
C-30 X X X 0 A
C-31 X X X 10 A
DC Amps
(with 50-turn Coil)
Save X X X STBY
C-32 X X X 60 mV @ 50 Hz
C-33 X X X 30 mV @ 50 Hz
C-34 X X X 60 mV @ 100 Hz
C-35 X X X 120 mV @ 200 Hz
C-36 X X X 180 mV @ 300 Hz
C-37 X X X 240 mV @ 400 Hz
C-38 X X X 300 mV @ 500 Hz
iFlex Current Probe
(Simulation)
Save X X X STBY