Clamp Meter
Calibration Adjustment
13
Table 4. Adjustment Procedure
Test
(Switch Position)
LCD
Reading
374 375 376 Calibrator Output
C-00 X X X 600 V @ 50 Hz
C-01 X X X 300 V @ 50 Hz
C-02 X X X 300 V @ 100 Hz
C-03 X X X 300 V @ 200 Hz
C-04 X X X 300 V @ 300 Hz
C-05 X X X 300 V @ 400 Hz
C-06 X X X 300 V @ 500 Hz
K
AC Volts
Save X X X STBY
C-07 X X X 0 V
C-08 X X X 600 V
C-09 X X X 0 V
C-10 X X X 0.5 V
L
DC Volts
Save X X X STBY
C-11 X X X 0 Ω
C-12 X X X 600 Ω
C-13 X X X 660 Ω
C-14 X X X 6000 Ω
C-15 - X X 6600 Ω
C-16 - X X 60000 Ω
C-17 X X X 0.1 μF
C-18 X X X 0.5 μF
C-19 X X X 1.5 μF
C-20 X X X 110 μF
C-21 X X X 500 μF
C-22 X X X 1000 μF
Ohms/Capacitance
Save X X X STBY