Performance Verification
4.4 General Instructions 4
4-5
4.4 General Instructions
Follow these general instructions for all tests:
It is recommended to remove the SD memory card from the Analyzer during the
performance test to avoid that measuring data is written on it during the test. The SD
card is located behind the battery door.
For all tests, power the Analyzer with its power adapter. The battery pack must be
installed and charged sufficiently.
Allow the Calibrator to satisfy its specified warm-up period.
For each test point , wait for the Calibrator to settle.
Allow the Analyzer a minimum of 30 minutes to warm up.
One division on the LCD consists of 25 pixels (1 pixel = 0.04 division).
4.5 Operating Instructions
Resetting the Analyzer
Proceed as follows:
Press
to turn the Analyzer off.
Press and hold
SAVE
SCREEN
.
Press and release
to turn the Analyzer on.
Wait until the Analyzer has beeped twice, and then release
SAVE
SCREEN
. When the
Analyzer has beeped twice, the RESET was successful. This reset is used as start
condition for the tests.
If the SD card has been removed, press
F5
- CLOSE.
Next press
F1
- CANCEL.
4.6 Display and Backlight Test
Proceed as follows to test the display and the backlight:
1. Press
to turn the Analyzer on. Press
F5
- CLOSE and
F1
to remove the
Language selection screen.
2. Disconnect the Power Adapter, and verify that the backlight is dimmed.
3. Connect the Power Adapter and verify that the backlight brightness increases.
4.7. Verification of Current Inputs
4.7.1 Preparation
Proceed as follows:
Reset the Analyzer and press
F1
- CANCEL.
Press
SETUP
. Then press F4 – MANUAL SETUP and use arrow keys and ENTER
key to set the Analyzer to 3-phase WYE, 60 Hz, 120 V.

Table of Contents

Other manuals for Fluke 435-II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 435-II and is the answer not in the manual?

Fluke 435-II Specifications

General IconGeneral
CategoryPower Quality Analyzer
Frequency Range42.5 Hz to 69 Hz
Harmonics MeasurementUp to 50th harmonic
Battery LifeUp to 8 hours
Safety RatingCAT IV 600 V / CAT III 1000 V
Inrush CurrentYes
FlickerYes
Mains SignalingYes
Motor AnalysisYes
Data LoggingYes
Voltage Measurement RangeUp to 1000 V
Current Measurement RangeDepends on current clamp used
DisplayColor LCD
PowerRechargeable Li-ion battery