EasyManua.ls Logo

Fluke 435-II - Verification of Voltage Inputs in 400 V Range

Fluke 435-II
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Verification
4.8 Voltage Inputs 4
4-13
Table 4-8. Accuracy Check of Voltage Channel N (Neutral)
Set Calibrator to Readout at Voltage Channel N
115 V, 50 Hz, OPR Fluke 434-II: 113.8 ... 116.2 V
Fluke 435-II /437-II: 114.77 ... 115.23 V
230 V, 50 Hz, OPR Fluke 434-II: 228.8 ... 231.2 V
Fluke 435-II /437-II: 229.77 ... 230.23 V
460 V, 50 Hz, OPR Fluke 434-II: 458.8 ... 461.2 V
Fluke 435-II /437-II: 459.77 ... 460.23 V
15. Set the Calibrator to STBY.
Optional Test. Bandwidth Check of Channel N (*):
16. Set the Calibrator to 215 V, 50 Hz and OPR.
17. Now check the voltage channel N. Use the Calibrator’s field edit function
(AMPL/FREQ key) to adjust the Calibrator to an Analyzer readout of 215.0 V /
215.00 V.
18. Increase the frequency to 3 kHz.
Fluke 434-II: check for a readout of 204.2 V or more.
Fluke 435-II/437-II: check for a readout of 204.25 V or more.
19. Set the Calibrator to STBY.
4.8.4 Verification of voltage inputs in 400 V range
Accuracy Check of Channel A/L1, B/L2, C/L3:
1. Set the Analyzer to 400 V, 50 Hz nominal (SETUP menu, arrow and ENTER keys).
Press MENU, select Volts/Amps/Hertz, press F5 - OK.
2. To check the A/L1, B/L2, C/L3 inputs: connect the N input to Ground (See Figure 4-
3).
3. Set the Calibrator to 50 Hz, 200 V and then to OPR.
4. Fluke 434-II: check for a voltage readout V rms between 198.0 ... 202.0 V.
Fluke 435-II/437-II: check for a voltage readout V rms between 199.60 ... 200.40 V.
5. Check the 400 V range according to the table below.

Table of Contents

Other manuals for Fluke 435-II

Related product manuals