TI mV Dial Voltage Divider Tolerance
Setting Setting Limits
1000 0.000999X
2 V or less
700 0.000699X ”
500 0.000499X ”
200 0.000199X ”
100 0.000099X ”
00*
3 V or less
*Connect the null detector II common lead to the low terminal of the TI, and wait until the thermal emf
caused by the heat from the operator’s hands has subsided.
4.3.7 Set the null detectors operate–zero switch–to zero.
4.3.8 Set the voltage sources for minimum output, and disconnect the test configuration.
4.4 10 VOLT DC DECADE TEST
4.4.1 Connect equipment as shown in Figure 4.
Figure 4. DC Volts Decade Test Configuration
2.4.3
DC VOLTAGE
CALIBRATOR
GUARD SENS OUT
+
–
TI
NULL DETECTOR
2.4.1
(BATTERY OPERATED)
IN
COM
GUARD
INPUT
OUTPUT
1.1 1.0 LO HI LO
VOLTAGE DIVIDER
2.5
4.4.2 Set the TI voltage output control(s) to obtain a 10V dc output and the voltage divider control(s) to
0.999999X.
NAVAIR 17–20AQ–70
9