TEST INST (S)
CALIBRATION CHECKLIST
PROC. NO.
Fluke 515A Portable Calibrator
MFG. MODEL
SER. NO.
PROCEDURE
STEP
NO.
(1)
FUNCTION TESTED
(2)
NOMINAL
(3)
MEASURED VALUES
FIRST RUN
(4)
SECOND RUN
(5)
OUT
OF
TOL (6)
CALIBRATION TOLERANCES
(7)
NA 17–20AQ–70
Page 1 of 2
4.1 Line Regulation Test
4.1.2 Line voltage 105 –– ck ( ) +
10 mV or less
4.1.4 ” 125 –– ck ( ) ”
4.2 DC Voltage Range Test
TI Range (mV) (mV)
4.2.7.2 100V 0 +
3 mV or less
”(mV) (mV)
4.2.7.2 10V 0 +
300 mV or less
1V 0 +
30 ”
4.3 DC Microvolts Output and Linearity Test
TI mV Divider (mV) (mV)
4.3.6 1000 0.000999X 0 +
2 mV or less
” 700 0.000699X 0 ”
” 500 0.000499X 0 ”
” 200 0.000199X 0 ”
” 100 0.000099X 0 ”
” 0 0.0000000 0 +
3 mV ”
4.4 10 Volts DC Decade Test
4.4.3 10 (1) 0.999999X –– ck ( ) Ref Null
TI Output Ratio (mV) (mV)
4.4.4 10 (0.9) 0.899999(X) 0 +
270 mV or less
” 10 (0.8) 0.799999(X) 0 +
240 ”
” 10 (0.7) 0.699999(X) 0 +
210 ”
” 10 (0.6) 0.599999(X) 0 +
180 ”
” 10 (0.5) 0.499999(X) 0 +
150 ”
” 10 (0.4) 0.399999(X) 0 +
120 ”
” 10 (0.3) 0.299999(X) 0 +
90 ”
” 10 (0.2) 0.199999(X) 0 +
60 ”
” 10 (0.1) 0.099999(X) 0 +
30 ”
NAVAIR 17–20AQ–70