TEST INST (S)
CALIBRATION CHECKLIST
PROC. NO.
Fluke 515A Portable Calibrator
MFG. MODEL
SER. NO.
PROCEDURE
STEP
NO.
(1)
FUNCTION TESTED
(2)
NOMINAL
(3)
MEASURED VALUES
FIRST RUN
(4)
SECOND RUN
(5)
OUT
OF
TOL (6)
CALIBRATION TOLERANCES
(7)
NA 17–20AQ–70
Page 2 of 2
4.5 AC Frequency Response Test up to 50 KHz
TI Range Freq (DC Volts) (DC Volts) (DC Volts)
4.5.8 10V 400 Hz 10.000 9.996 to 10.004
4.5.9 10V 4 kHz 10.000
10V 50 kHz 10.000 9.990 to 10.010
4.6 AC Voltage Test
TI Output
Range Freq (%) (%)
4.6.2.2 1V 400 Hz 0 +
0.058% or less
100V 0 +
0.068% or less
4.7 Frequency Test
TI Output
Volts Freq (Hz) (Hz) (Hz)
4.7.2 10 400 Hz 400 396 to 404
10 4 kHz 4000 3960 to 4040
10 50 kHz 50000 47500 to 52500
4.8 Resistance Test
TI Output
(W)(W)(mW)
4.8.11 0 (Residual) –– +
150 or less
4.8.12 10 –– +
(6 + residual) or less
100 –– +
(60 + residual) or less
(kW)(W)(mW)
4.8.12 1 –– +
(150 + residual) or less
”(kW)(W)(W)
4.8.12 10k –– +
(1.5 + residual) or less
100k –– +
(15 + residual) or less
(MW)(W)(W)
4.8.12 1 –– +
(150 + residual) or less
”10 +
(7500 + residual) or less
NAVAIR 17–20AQ–70