EasyManua.ls Logo

Fluke 5520A - Page 20

Fluke 5520A
38 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5520A
Getting Started
14
AC Voltage (Sine Wave) Specifications (cont)
AUX (Auxiliary Output) [dual output mode only] [1]
Absolute Uncertainty,
tcal ± 5 °C
± (% of output + µV)
Max Distortion
and Noise
10 Hz to
100 kHz
Bandwidth
Range Frequency 90 days 1 year
Res-
olution
Max
Burden
± (% output +
floor)
10 Hz to 20 Hz 0.15 + 370 0.2 + 370
0.2 + 200 µV
20 Hz to 45 Hz 0.08 + 370 0.1 + 370 0.06 + 200 µV
45 Hz to 1 kHz 0.08 + 370 0.1 + 370 0.08 + 200 µV
1 kHz to 5 kHz 0.15 + 450 0.2 + 450 0.3 + 200 µV
5 kHz to 10 kHz 0.3 + 450 0.4 + 450 0.6 + 200 µV
10 mV to
329.999 mV
10 kHz to 30 kHz 4.0 + 900 5.0 + 900
1 µV5 mA
1 + 200 µV
10 Hz to 20 Hz 0.15 + 450 0.2 + 450
0.2 + 200 µV
20 Hz to 45 Hz 0.08 + 450 0.1 + 450 0.06 + 200 µV
45 Hz to 1 kHz 0.07 + 450 0.09 + 450 0.08 + 200 µV
1 kHz to 5 kHz 0.15 + 1400 0.2 + 1400 0.3 + 200 µV
5 kHz to 10 kHz 0.3 + 1400 0.4 + 1400 0.6 + 200 µV
0.33 V to
3.29999 V
10 kHz to 30 kHz 4.0 + 2800 5.0 + 2800
10 µV5 mA
1 + 200 µV
10 Hz to 20 Hz 0.15 + 450 0.2 + 450
0.2 + 200 µV
20 Hz to 45 Hz 0.08 + 450 0.1 + 450 0.06 + 200 µV
45 Hz to 1 kHz 0.07 + 450 0.09 + 450 0.08 + 200 µV
1 kHz to 5 kHz 0.15 + 1400 0.2 + 1400 0.3 + 200 µV
3.3 V to 5 V
5 kHz to 10 kHz 0.3 +1400 0.4 + 1400
100 µV5 mA
0.6 + 200 µV
[1] There are two channels of voltage output. The maximum frequency of the dual output is 30 kHz.
Note
Remote sensing is not provided. Output resistance is < 5 m for outputs 0.33 V. The AUX output resistance is
< 1. The maximum load capacitance is 500 pF, subject to the maximum burden current limits

Other manuals for Fluke 5520A

Related product manuals