AC Voltage, Square Wave Characteristics
Risetime
@ 1 kHz
Typical
Settling
Time
@ 1 kHz
Typical
Overshoot
@ 1 kHz
TypicalDuty Cycle RangeDuty Cycle Uncertainty
<1 µs<10 µs to 1%
of final value
<2%1% to 99%, <3.3 V p-p,
0.01 Hz to 100 kHz
± (0.8% of period +100 ns)
± (0.02% of period + 100
ns), 50% duty cycle
± (0.05% of period + 100
ns), other duty cycles from
10% to 90%
AC Voltage, Triangle Wave Characteristics (typical)
Linearity to 1 kHzAberrations
0.3% of p-p value, from 10% to 90% point<1% of p-p value, with amplitude >50% of range
AC Current (Sine Wave) Extended Bandwidth Specifications
RangeFrequency
1-Year Absolute Uncertainty
tcal ± 5 °C
± (% of output + % of range)
Max
Current
Resolution
All current ranges, <330 mA0.01 Hz to 10 Hz5.0 + 0.52 digits