5730A
Operators Manual
6-18
Table 6-5. Commands (cont.)
CAL_CLST?
Description
Sequential command.
Returns a list of names values of a particular group of calibration constants.
Parameter
CAL (Returns active calibration constants)
CHECK (Returns calibration check constants)
PREV (Returns previous calibration constants)
Response
(String)
"<EOL>
<total number of constants> <EOL>
<name> <value> <EOL>
<name> <value> <EOL>
(cont)
"
Example
CAL_CLST?, CAL Could return:
"
424
D3P, 3.9817876E+02
(Continues for about 400 pairs of names and values.)
CAL_CONF
Description
Sequential command.
Sets the calibration specifications to a confidence level of either 99% or 95%.
Paramete
r
CONF95 or CONF99: Sets 95% or 99% specifications.
CAL_CONF?
Description
Sequential command.
Returns the current calibration confidence level.
Paramete
r
None
Response CONF95
(
indicates 95% specifications
)
a or CONF99
(
indicates 99% specifications
)
CAL_CONST?
Description
Sequential command.
Returns the value of a particular calibration constant from a particular
g
roup of constants.
Parameter
1. CAL (From active calibration constants)
CHECK (From calibration check constants)
PREV (From previous calibration constants)
2. S
y
mbolic name of the calibration constant desired
(
see Appendix D
)
.
Response
(
Float
)
The value of the constant
Example
CAL_CONST? CHECK, KV6
Could return: 6.5000010E+00
CAL_COUNT?
Description
Sequential command. Returns the number of calibration constants that have been saved
at the end of a calibration procedure
except for DC Zeros
Response inte
g
e
r
Parameter None
Example
CAL_COUNT?
Returns 34 if calibration constants have been saved 34 times.