Front-Panel Operation
Connect the Calibrator to a UUT 4
4-7
Figure 4-4B shows a meter connected in a two-wire connection with the two-wire
compensation circuitry turned off. For low resistances where uncompensated lead
resistances are significant, use the two-wire compensation circuit and the connection in
Figure 4-4C or 4-4D. Use the connection in Figure 4-4C to calibrate the meter referenced
at its terminals. Use the connection in Figure 4-4D to calibrate the meter referenced at the
end of its test leads.
Cable Connection Instructions
To connect the 5730A Calibrator to a UUT:
1. If the Calibrator is powered on, push , or . Either action sets the
Calibrator to 0 mV in standby.
2. Determine the appropriate figure from Table 4-2, and refer to the preceding text and
Table 4-1 to make appropriate connections to the UUT.
Table 4-2. UUT Connection Figures
5730A Output Figure
DC Voltage (including 5725A boost) 4-1
AC voltage ≤10 kHz 4-1
AC voltage >10 kHz 4-2
AC current ≤2A, ≤10 kHz 4-3
Resistance 4-4
Wideband AC voltage
(option 5730A/03 or 5730A/05)
4-5
5725A amplified output (current only) 4-6