5730A
Operators Manual
4-12
UUT
COMP
COMP
Calibrator
UUT
COMP
COMP
Calibrator
UUT
Calibrator
Calibrator
UUT
Caution:
Use connections with
exposed plug tips for
the ohms function only.
NC
NC
NC
C
D
2-WIRE
COMP
ON
External
Sense
ON
External
Guard
OFF
2-WIRE
COMP
ON
External
Sense
OFF
External
Guard
OFF
hhp015.eps
Figure 4-4. UUT Connections: Resistance (cont.)