5730A
Operators Manual
4-18
9. Enter a frequency with the numeric keypad followed by or if necessary. The
display now shows the amplitude and frequency of the entry. If an entry error is
made, use to clear the most recently entered digit, or push to clear the
display. Reenter the value. The display shown below assumes an entry of 2.8 kHz.
+/- 0.00050mV (1 year) 213 days since calibration
3/14/2013 10:52am
External
Guard
OFF
Wideband
OFF
Setup
Menu
Standby
mV
0.00000
Auto Range 220 mV
19V 2.8kHz
External
Sense
OFF
hhp115.eps
10. Push . The Calibrator clears the entry from the entry bar toward the bottom
and shows in the larger area above. No voltage is available at the output terminals
until is pushed.
11. Push to activate the Calibrator output. The UUT will now respond to the
applied voltage.
In AC voltage output, the selections available at the bottom are:
External Sense
External Guard
Boost
Wideband
Setup Menu
In addition, the Phase Control and Scale functions are available. The output range
is always automatically chosen and cannot be locked in ac volts.
Note
The Calibrator stays in the ac function after a non-zero frequency is
entered. To change back to dc, enter a frequency of 0 Hz or a signed
voltage.