Process Calibrators
Calibration
61
712 Calibration (V1.2 through V1.9)
1. Hold down and while turning the power on.
2. Hold the two keys until CAL is displayed, then release the two keys. The display
should read:
INPUT 15 Ω
3. Connect the two test leads from the NORMAL jacks of the 5520A to the INPUT
jacks (two middle jacks) on the 712 Calibrator (black to black, red to red).
4. Connect two additional test leads, also from the NORMAL jacks of the 5520A, to the
712 Calibrator (black wire to the 712 red jack labeled [3W NC] and red wire to the
red jack labeled [4W NC]). All four input jacks of the 712 should now be configured
in a 4-wire ohm measurement.
5. Set the Fluke 5520A to 15Ω and push O.
6. Push . The display should read:
INPUT ----- Ω,
then change to:
INPUT 350 Ω
7. Set the Fluke 5520A to 350 Ω.
8. Push . The display should read:
INPUT ----- Ω
then change to:
INPUT 500 Ω.
9. Set the Fluke 5520A to 500 Ω.
10. Push . The display should read:
INPUT ----- Ω
then change to:
INPUT 3200 Ω
11. Set the Fluke 5520A to 3200 Ω.
12. Push . The display should read:
INPUT ----- Ω.
then change to:
OUTPUT 17-50 Ω.
13. Using four test leads, connect the four jacks of Fluke 8508A (Ω Sense 'Hi & Lo' and
Input 'Hi & Lo') to the two middle jacks of the 712 Calibrator (black to black; red to
red). This puts the Fluke 8508A in a 4-wire configuration.
14. Set the 3458A to the 1000 Ω range.
15. Push and until the 712 display matches the Fluke 8508A.
16. Push . The display should read:
OUTPUT ----- Ω
then change to:
OUTPUT 350.00 0.
17. Push and until the 712 display matches the Fluke 8508A.
w ww . . co m
information@itm.com1.800.561.8187

Table of Contents

Other manuals for Fluke 715

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 715 and is the answer not in the manual?

Fluke 715 Specifications

General IconGeneral
Battery voltage9 V
Battery technologyAlkaline
Display typeLCD
Display digits5 digits
Drop proof up to1 m
Operating altitude0 - 3000 m
Operating vibration2 G
Storage temperature (T-T)-40 - 60 °C
Operating temperature (T-T)-10 - 55 °C
Operating relative humidity (H-H)0 - 95 %
Holster includedYes
Weight and Dimensions IconWeight and Dimensions
Depth32 mm
Width87 mm
Height187 mm
Weight330 g

Related product manuals