Process Calibrators
Calibration
81
718Ex Calibration
mA Measure
1. Hold down both and on the 718Ex Pressure Calibrator. Push and release .
2. The display shows CAL xxx, as long as and are pressed. Where xxx is a
number, indicating the number of times the 718Ex has been through the calibration
adjustment. When and are released,
PASS is displayed. The password is
required before you can enter the calibration mode. The password is 817. Enter the
most significant digit first. The password appears below
PASS. Use to increase
the displayed number, to decrease it. When the required digit appears, push
to proceed to the next digit, or the password check. If the password entry was correct,
the 718Ex will go to the CAL mode. If the password is incorrect,
BAD PASS shows
on the display and then the 718Ex returns to the normal operation mode. In the CAL
mode, the 718Ex should display:
0.000 mA
in the upper display and
CAL 0 mA
on the lower line of the display.
The upper display shows an uncalibrated reading. This reading is approximately
±10 % of full scale of the applied reading. When is pressed, the calibration value
is calculated relative to this reading. The lower display indicates what function and
point is being calibrated.
3. Connect the test leads from the AUX jacks of the Fluke 5520A to the mA jacks on
the 718Ex Calibrator (black to black and red to red).
4. Set the Fluke 5520A to 0.00000 mA and push O.
5. When 0.0 mA is sourced from the 5520A the upper display indicates about
"0.000 mA". When is pushed, the unit establishes the calibration constant for
that point and the lower display changes to "CAL 24 mA". The upper display still
reflects the applied current so it does not change until after step 6 when 24 mA is
applied."
6. Set the Fluke 5520A to 24.0000 mA.
7. The upper display reads approximately 24.000 mA. Push on the 718Ex
Calibrator. CHEC appears on the lower display.
8. The CHEC mode allows for a quick check of the calibration constants that have just
been stored. Changes to the applied current are reflected in the display. Push to
continue. The lower display changes to:
rAnGE
w ww . . co m
information@itm.com1.800.561.8187

Table of Contents

Other manuals for Fluke 715

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 715 and is the answer not in the manual?

Fluke 715 Specifications

General IconGeneral
Battery voltage9 V
Battery technologyAlkaline
Display typeLCD
Display digits5 digits
Drop proof up to1 m
Operating altitude0 - 3000 m
Operating vibration2 G
Storage temperature (T-T)-40 - 60 °C
Operating temperature (T-T)-10 - 55 °C
Operating relative humidity (H-H)0 - 95 %
Holster includedYes
Weight and Dimensions IconWeight and Dimensions
Depth32 mm
Width87 mm
Height187 mm
Weight330 g

Related product manuals