Zeroing with Absolute Pressure Modules
9
Zeroing with Absolute Pressure Modules
For zeroing, adjust the Calibrator to read a known
pressure. This can be barometric pressure, if it is
accurately known, for all but the 700PA3 module. An
accurate pressure standard can also apply a pressure
within range for any Absolute Pressure Module. Adjust the
Calibrator reading as follows:
1. Press and hold
Z
.
2. Press
a(Z) to increase or
h
(Y) to decrease
the Calibrator reading to equal the applied pressure.
3. Release
Z
to exit the zeroing procedure.
Press the
U button to convert to any convenient