EasyManua.ls Logo

Fluke 718 1G - Specifications; Pressure Sensor Input; Pressure Module Input; DC Ma Input

Fluke 718 1G
31 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Pressure Calibrator
Specifications
25
Specifications
Specifications are based on a one year calibration cycle
and apply for ambient temperature from +18 °C to +28 °C
unless stated otherwise. “Counts” are the number of
increments or decrements of the least significant digit.
Pressure Sensor Input
Model Range Accuracy
Max Non-
destructive
Pressure
1G -1 to 1 PSI
(-7 to 8 kPa)
5 PSI
(34.5 kPa)
30G -12 to 30 PSI (-
83 to 207 kPa)
60 PSI
(413 kPa)
100G -12 to 100 PSI
(-83 to 690
kPa)
200 PSI
(1.4 mPa)
300G -12 to 300 PSI
(-83 to 2068
kPa)
± 0.05 % of
Range
375 PSI
(2.6 mPa)
Temperature coefficient: 0.01 % of range per
°
C for
temperature ranges -10
°
C to 18
°
C and 28
°
C to 55
°
C
Pressure Module Input
Range Resolution Accuracy
(determined by Pressure Module)
DC mA Input
Range Resolution Accuracy, ±(% of
Reading + Counts)
24 mA 0.001 mA 0.015 + 2
Fuseless overload protection
Temperature coefficient: 0.005 % of range per
°
C for
temperature ranges -10
°
C to 18
°
C and 28
°
C to
55
°
C
Loop Supply
24 V dc nominal
Shop for Fluke products online at:
1.888.610.7664
www.MyFlukeStore.com

Other manuals for Fluke 718 1G

Related product manuals