Users Manual
56
MOTOR_OFF
Turns off the motor.
MOTOR?
Returns the state of the motor.
For example:
If the motor was on, MOTOR? returns ON.
OHMS?
Returns the raw ohm value from the RTD.
For example:
If when a P100-385 is measured at 0 °C, OHMS? returns
1.000000E+02, OHM.
OUT
This command also switches the Product into mA output
mode. A number and a unit must be entered after the
command.
For example:
OUT 5 MA sets the current output at 5 mA
OUT?
Returns the output of the Product.
With the above example, OUT? returns 5.000000E-03, A
PRES_UNIT
Used to set the pressure unit for the indicated display.
For example:
To set the pressure unit to psi on the lower display, send
PRES_UNIT LOWER, PSI.
PRES_UNIT?
Returns the pressure unit used when pressure is
measured for each of the 3 displays.
PUMP_LIMIT
Sets the approximate pressure in psi at which the pump
will turn off.
For example:
PUMP_LIMIT 50 sets the approximate value that the
pump will shut off at to 50 psi