79/26
Calibration Information
6
Recommended Equipment
Test equipment recommended for the performance tests and calibration is listed in Table 2. If the
recommended models are not available, instruments with equivalent specifications may be used.
Table 2. Required Equipment
Equipment Minimum Specifications Recommended Model
DMM Calibrator DC Voltage: 0-600V Fluke 5100B, 5101B,
plus Transconductance Accuracy: .05 % 5102B, 5700A) + 5220A or
or Power Amplifier AC Voltage: 0 to 1000 V
Accuracy: 0.2 %
Frequency: 45 Hz to 20 kHz
DC mA: 0 to 35 mA
DCA: 0 to 10 A
Accuracy: 0.1 %
AC mA: 0 to 35 mA
ACA: 0 to 10 A
Accuracy: 0.3 %
Frequency: 45 Hz to 1 kHz
Fluke 5500A Multi-Product
Calibrator, 5700A + 5725A
Function Generator Sinewave voltage: 0 to 1 V rms
Frequency: 1 Hz to 20 kHz
Frequency Accuracy: 002 %
5500A Multi-Product Calibrator
or equivalent
Decade Resistor Resistance 0 to 35 M:
Accuracy: .05 %
Fluke 5500A Multi-Product
Calibrator or equivalent
Decade Capacitor Capacitance: 100 pF to 1.1 µF
Accuracy: 0.5 %
Fluke 5500A Multi-Product
Calibrator or equivalent

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 79 III and is the answer not in the manual?

Fluke 79 III Specifications

General IconGeneral
CategoryMultimeter
Display TypeDigital
True RMSNo
Voltage DC1000V
Voltage AC750V
Current DC10A
Current AC10A
Resistance40MΩ
ContinuityYes
Diode TestYes
Battery Type9V
Resistance Range400Ω to 40MΩ
Display4000 count LCD
DC Voltage Range400 mV, 4 V, 40 V, 400 V, 1000 V
AC Voltage Range400 mV, 4 V, 40 V, 400 V, 750 V
DC Current Range10μA to 10A
AC Current Range10μA to 10A

Related product manuals