7. TEMPERATURE PROBE (80T-150)
6
-
8. lntroduction
.
"
6
-
9. The 80T
-
150 Temperature Probe converts the
8020A into a direct
-
reading (1 mV dc/O) OC or OF
thermometer. It is ideally suited for surface, ambient and
liquid measurement, and lends itself easily to a wide
range of design, troubleshooting and evaluation
applications. A rugged, fast
-
responding probe
-
tip with a
350V dc standoff makes the 80T
-
150 one of the most
versatile and easy
-
to
-
use temperature probes available.
,
,
6
-
10. Specifications
Range(OCI0F)
...
.-50°C to +150°C or
field selectable by
-58OF to 302OF
internal jumpers)
Accuracy
.........
*l°C (1.8OF) from "OC
to 100°C, decreasing
linearly to
f
3OC (5.4OF)
at -50°C and +150°C
Resolution
.......
.O. 1 C on 200 mV range
Voltage Standoff.
.
.350V dc or peak ac
Power
............
Internal disposable battery;
1,000 hours of continuous use.
6
-
11. CURRENT TRANSFORMER (801
-
600)
6
-
12. lntroduction
6
-
13. The Model 801-600 extends the maximum 2A ac
current measuring capability of the
8020A up to
a
maximum of 600 amps. A clamp
-
on transformer
designed into the probe allows measurements to be made
without breaking the circuit under test. In use, the current
carrying conductor being measured serves as the
transformer's primary while the
801-600 serves as the
secondary. Because of a high efficiency, quadrature type
of winding, wire size and location of the conductor within
the transformer jaws do not affect accuracy of the current
measurement.
6
-
14. Specifications
Range
...........
.2 to 600A ac
Accuracy
.........
f
3%
Frequency
.......
.30 Hz to 1 kHz
Response
Division Ratio
....
.1000: 1
Insulation
Insulation
.......
.5 kV
Maximum
.......
.2
-
inch diameter
Conductor Size
6
-
2
6
-
15. HlGH VOLTAGE PROBE (80K-40)
6
-
16. lntroduction
6
-
17. The Model 80K
-
40 extends the voltage measure
-
ment capability of the 8020A up to 40 kV. Internally, the
probe contains a special 1000:
1 resistive divider. Metal-
film resistor with matched temperature coefficients
comprise the divider, and provide the probe with its
excellent accuracy and stability characteristics. Also, an
unusually high input impedance (1000
Ma) minimizes
circuit loading, and thereby contributes to measurement
accuracy.
6
-
18. Specifications
....
Voltage Range
.I
kV to 40kV dc or peak ac,
28kV rms ac.
Input Resistance
..
.I000 MR
Division Ratio
....
.1000:1
Accuracy DC
Overall Accuracy: 20
kV to 30 kV
f
2% (calibrated
at 25
kV).
Upper Limit: Changes linearly from 2% at 30 kV to 4%
at
40
kV.
Lower Limit: Changes linearly from 2% at 20 kV to 4%
at
1
kV.
Accuracy AC
......
f
5% at 60 Hz
(Overall)
6
-
19. HlGH FREQUENCY PROBE (81RF)
6
-
20. lntroduction
6
-
21. The 81RF Probe extends the frequency range of
the
8020A's voltage measurements capability to include
100 kHz to 100 MHz inputs from 0.25 to 30V rms. It
operates in conjunction with the
8020A dc voltage
ranges, and provides a dc output that is calibrated to be
equivalent to the rms value of a
sinewave input.
6
-
22. Specifications
Frequency
.......
.f
1 dB from 100 kHz to
Response 100 MHz
Extended
........
.Useful for relative reading
Frequency from 20 kHz to 250 MHz.
Response
Response
........
.Responds to peak value of input;
calibrated to read rms value of
a
sinewave
Voltage Range
....
.0.25 to 30V rms
Maximum DC
....
.350V dc
Input
Input Impedance
..
.12Ma shunted by 15 pF